Last N = previous measurement reading
F average = (1-x)* F4 + x* F5
N(R/h) = [n*F1 + dead time correction {n*F average}/300
The following graph illustrates the ranges over which F2, F3 and F average are used as the DRM-1’s dead time
correction factor:
9. Calibration Summary for the DRM-1D/2D
9.1 Preface
9.1.1 Calibration Factors
The DRM-1D/2D requires only one calibration point which requires only one change to Factor 1 (F1) in the
M
eter Calibration software. Once the F1 factor has been set performed, linearity verification should be
performed.
The DRM-1D/2D can be calibrated via the RS232 communication port with a NULL modem adapter.
Switch to the ON position dipswitches 1&2 for RS232 and the calibration will be performed by using the Meter
Calibration software. Refer to section 10.1 for instructional use of the calibration software
9.1.2 Calibration Ranges
F1 is recommended to be set at 0.1 to 0.2 mR/h.
Dead Time