MGP WRM2 Especificações Página 10

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 9
TelePole WR/ Operating & Maintenance Manual Page 10
3.3.3. Reading reset:
Exposure rate - While in Exposure rate mode, a long press on the RESET push-button
enters the meter into the FREEZE mode (for peak measuring). An additional long press
performs a momentary reset of the Exposure rate readings.
The reset function provides a rapid means of discharging the Display reading and enables
accurate measurement of low level Exposure rates.
3.3.4. Audible announcements:
When the meter is turned on, the radiation field intensity is observed by the click rate. A
long press on the MODE push-button toggles the click on and off.
Each push-button press is accompanied by an audible beep.
In case of threshold alarm or failure alarm the audible alarm is activated. To mute the
audible alarms press the RESET push-button.
High Range
Geiger
Low Range
Overflow
alarm
Field
800 mR/h
1000 R/h
ZP 1301
ZP 1201
600 mR/h
0.01 mR/h
High Range
Geiger Type Optional 070301 Firmware
Low Range
Overflow
alarm
Field
2500 mR/h
1000 R/h
ZP 1301
ZP 1201
2000 mR/h
0.01 mR/h